Integritest? 4 Series Automated Filter Integrity Test Instruments
如果您對該產品感興趣的話,可以
產品名稱: Integritest? 4 Series Automated Filter Integrity Test Instruments
產品型號: XIT4S0001
產品展商: MILLIPORE
產品文檔: 無相關文檔
簡單介紹
Integritest? 4 Series Automated Filter Integrity Test Instruments
Integritest? 4 Series Automated Filter Integrity Test Instruments
的詳細介紹
Integritest® 4 Series Automated Filter Integrity Test Instruments
Description
The Integritest 4 instrument is an easy-to-use, portable, fully automated integrity test system. Available in either a standalone or networkable configuration, the Integritest 4 instrument offers reliable, repeatable integrity test data. The intuitive touch screen user interface streamlines the test process and the accelerated testing capability provides users with more time for processing. On-site calibration support and diagnostic capability minimizes downtime, saving time and money. Leveraging a proven test algorithm, the Integritest 4 instrument performs bubble point, diffusion, enhanced bubble point, and HydroCorrSM (Millipore water-based test for hydrophobic filters) tests on a wide range of filters including disks, cartridges and TFF filters, virus, and asymmetric membrane filters.
The rugged, ergonomic design of the system enables users to easily carry the Integritest 4 instrument to any site. The standalone configuration enables users to print the results on the onboard printer.
The Integritest 4N networkable configuration enables users to print results directly to any networked printer via Ethernet connection. The Integritest 4N instrument is 21 CFR Part 11 compliance ready for electronic records and electronic signatures, improving productivity and facilitating record keeping.
- Portable, easy to use
- Reliable, accurate measurements
- Fast testing improves productivity
- Network model is 21 CFR Part 11 compliance ready
- Available in multiple languages
Power Requirements |
Voltage |
90–264 V / 50–60 Hz * |
Current rating |
3.5 A |
Compressed Gas |
Inlet pressure |
Clean, dry air or nitrogen source of 2.4 – 8.2 bar (35 – 120 psi) at least 1 bar (15 psi) greater than the highest test pressure |
Operating pressure range, bar (psig) |
0.34 – 7.03 (5 – 102) |
Environmental |
Storage temperature, °C |
-20–80 |
Operating temperature, °C |
1–40 |
Humidity |
5 to 95%, non-condensing |
Nominal Dimensions |
Height, cm (in.) |
37.2 (14.6) |
Width, cm (in.) |
41.6 (16.4) |
Depth, cm (in.) |
25.4 (10.0) |
Weight, kg (lb) |
<15 (32) |
Computer |
Software operating system |
Windows XP embedded |
Port 1 |
5-pin DIN connector for service access |
Port 2 |
PS/2 Barcode Reader Input |
Port 3* |
RJ45 for Ethernet connection |
Test Accuracy (at standard, stable conditions) |
Test Type |
Range |
Accuracy |
Diffusion (Standard and Accelerated) |
Flow rates <20 mL/min |
≤1mL/min |
Flow rates ≥20 mL/min (water) |
≤5% |
Flow rates ≥20 mL/min (alcohol) |
≤10% |
30 Second Diffusion* |
Flow rates <20 mL/min |
≤2 mL/min |
Flow rates ≥20 mL/min |
≤15% /–10% |
Bubble Point (BP) Standard |
345 – 6205 mbar (5 – 90 psig) |
≤68.9 mbar (1 psig) |
Bubble Point Asymmetric |
345 – 6205 mbar (5 – 90 psig) |
<344.7 mbar (5 psig) |
Accelerated BP Hydrophilic PVDF |
345 – 6205 mbar (5 – 90 psig) |
≤137.9 mbar (2 psig) |
Accelerated BP Hydrophobic PTFE |
345 – 6205 mbar (5 – 90 psig) |
≤206.8 mbar (3 psig) |
HydroCorr |
Flow rates <0.4 mL/min |
≤0.02 mL/min |
Flow rates ≥0.4 mL/min |
≤5% |
Virus Diffusion |
Flow rates <20 mL/min |
≤2 mL/min |
Flow rates ≥20 mL/min |
≤10% |
Pressure Hold |
Pressure drops <69 mbar (1 psig) |
≤3.5 mbar (0.05 psi) |
Pressure drops ≥69 mbar (1 psig) |
≤5% |
Test Reproducibility (at standard, stable conditions) |
Coefficient of variation (CV) or standard deviation (sn–1) |
Test Type |
Range |
Reproducibility |
Diffusion* |
Flow rates <20 mL/min |
sn–1 ≤2 mL/min |
Flow rates ≥20 mL/min |
CV ≤10% |
All Bubble Point Tests |
345 – 6205 mbar (5 – 90 psig) |
CV ≤10% |
HydroCorr |
Flow rates <0.4 mL/min |
sn–1 ≤0.04 mL/min |
Flow rates ≥0.4 mL/min |
CV ≤10% |
Virus Diffusion |
Flow rates <20 mL/min |
sn–1 ≤2 mL/min |
Flow rates ≥20 mL/min |
CV ≤10% |
Pressure Hold |
Pressure drops <69 mbar (1 psi) |
sn–1 ≤6.9 mbar (0.1 psi) |
Pressure drops ≥69 mbar (1 psi) |
CV ≤10% |
Test Specifications |
Test Type |
Configuration Range |
Operating Range |
Diffusion* |
47 mm disks –
12 round x 30 in. systems
Capsules to TFF systems |
Diffusion rate: 1 – 600 mL/min |
Test pressure: 345–6550 mbar (5 – 95 psig) |
Bubble Point
(All Except Accelerated Hydrophobic) |
47 mm disks –
3 round x 30 in. systems |
Test pressure: 345–6205 mbar (5 – 90 psig) |
Accelerated Hydrophobic Bubble Point |
47 mm disks –
1 round x 30 in. systems |
Test pressure: 345–6205 mbar (5 – 90 psig) |
HydroCorr |
47 mm disks (Aervent-50) –
3 round x 30 in. systems |
Test pressure: 345–6205 mbar (5 – 90 psig) |
Flow rate: 0.01 – 10 mL/min |
Virus Diffusion* |
25 mm disks –
3 round x 30 in. systems |
Diffusion rate: 1 – 600 mL/min |
Test pressure: 345–6550 mbar (5 – 95 psig) |
Pressure Hold |
Up to 12 round x 30 in. systems |
Test pressure: 345–6550 mbar (5 – 95 psig) |